PROCEDURE
1. Take 5 calibrated thermometer and give them No. L1, L2, L3, L4, L5 as mentioned in annexure.
2. Keep five thermometer in test tubes at different location inside well of heating block as per location chart enclosed.
3. Operate Heating block as per S.O.P.
4. Compare temperature of each thermometer with respect to temperature adjusted one on heating block at every 15 minutes interval upto 1 hours.
5. Record the result as per annexure.
ACCEPTANCE CRITERIA
Temperature on display should be within 37°C + 1°C.
The results obtained at all location of heating block should be within 36 to 38°C.
Change the location by placing thermometer in other than previously selected location to cover maximum well during validation in a calendar year.
FREQUENCY
Every month